Real-Time Confocal Microscope (H3 Andor STORM Spinning disc)

Informationen

Name: Real-Time Confocal Microscope (H3 Andor STORM Spinning disc)
Einrichtung: Abteilung für Lichtmikroskopie (LMF)
Partner: Max-Planck-Institut für Molekulare Zellbiologie und Genetik (MPI-CBG)

Beschreibung

Category
Light Microscope
Overview
Fast, "real-time" confocal imaging with FRAPPA capability. Illumination equipment: TM-LED, Andor AMH, 405nm diode laser + 488 nm solid-state laser Coherent Sapphire + 561 nm solid-state laser + 640 nm solid-state laser.
Suitable for confocal imaging, photobleaching and -activation of DAPI,EGFP, mCherry, Cy5 with high temporal and spatial resolution
Features
Inverted stand
motorized with PFS (hardware autofocus)
1.5 x Optovar
MCL Piezo 100um travel range
motorized XY stage
FRAPPA coupled in via backport
Objectives
Nikon Plan-Apochromat 10x 0.45 DIC
Nikon Plan-Fluor 20x 0.75 Oil/Glyc/Water DIC
Nikon Plan-Apochromat 60x 1.2 W
Nikon Apo 100x/1.49 Oil DIC 0.13-0.20
Illumination
Transmitted Light (LED)
Fluorescence (Metal Halide)
Laser Diode 405 nm
Laser DPSS 488 nm
Laser DPSS 561 nm
Laser DPSS 640 nm
Detection
[[detection::Spinning disc scan head Yokogawa CSU-X1 (10000rpm)
manual:
dichromatic mirror:
Quad band T-405/488/568/647
physical pinhole radius: 25um
pinhole spacing: 250um
back-projected pinhole size: 0.4um
back-projected pinhole spacing: 4um (with 60x objective
1x optovar)
camera:
Andor iXon EM+ DU-897 BV back illuminated EMCCD
dexel size of EMCCD chip: 16um
image pixel sizes:
with 10x air obj.: 1.048um/px (1x optovar)
0.699 um/px (1.5x opt.)
with 20x Oil/Glyc/Water obj.: 0.524um/px (1x optovar)
0.349 um/px (1.5x opt.)
with 60x W obj.: 0.175um/px (1x opt.)
0.116um/px (1.5x opt.)
with 100x oil obj.: 0.105um/px (1x opt.)
0.0698um/px (1.5x opt.)]]
Reflectors
Rotr filter wheel:
position 0: 520/15
1: 628/40
2: 447/60
3: 607/36
4: dual 512/630
5: LP 664 RU
6: 525/30
7: 716/40
8: LP 488 razor edge
9: 617/73
Software
[[software::iQ 2.9 + iQ 3.4.1 (both with Python) iQ3 user guide
connections: COM1 - ????
COM2xx - lasers
COM3 - ???
COM4 - ???
COM5 - ???
COM6 - ???
AOTF - ALC401/501/601
DD A06/16]]
Incubation
Stage incubator (T+CO2)

Link zu weiteren Informationen

https://www.biodip.de/wiki/H3_Andor_STORM_Spinning_disc

Ansprechpartner

LMF
E-Mail:
Telefon:
+49 351 210-2345
Fax:
+49 351 210-1078